LOGiT Pressure and Temperature Logger
Supco 2 Channel Data Logger with External Sensors for Temperature and Pressure
The LOGiT pressure and temperature data logger is a 2 channel data logger with external sensors. The logger features alarms, fast sampling rate of 1 second, 21,500 data storage and minimum 18 month battery life.
Included with each Pressure and Temp Logger
>> Temperature Sensor
>> Pressure Sensor
Required with each LOGiT Data Logger
>> LOGiT Software (#LLS or #LLSU)
>> 9 Volt Battery
NOTE: NIST Calibration is available. Large Quantity Discounts Are Available!
Features Specifications
| Description |
Part Number |
Qty 1-9 |
Qty 10+ |
|
| LOGiT Pressure and Temperature Logger |
LPT
|
$256.00 |
call |
 |
| Software and Interface Kit - 1 Required |
| LOGiT PC Software and Serial Interface Cable Kit (Includes Software, Serial Cable and LINT Interface) |
LLS
|
$74.00 |
call |
 |
| LOGiT PC Software and USB Cable Kit (Includes Software, USB Cable and LINTU Interface) |
LLSU
|
$74.00 |
call |
 |
| Battery - Required |
| 9 Volt Lithium battery, 1200 mAH |
U9VL-FP
|
$10.00 |
call |
 |
| Standard NIST Calibration Service |
| Optional Accessories |
| USB Serial Adapter |
Cable-USB232
|
$39.00 |
$35.00 |
 |
| Keyspan USB 4-Port Serial Adapter |
USA-49WG
|
$149.00 |
call |
 |
LOGiT Temperature and Pressure Logger Features
- 2 Channels: External Sensors for Temperature and Pressure
- Store up to 21,500 Samples
- Sampling Rate of 1 Second to 18 Hours
- High and Low Alarms
- Single Point Offset Calibration through Software
- Requires LOGiT Software and 9 Volt Battery
LOGiT Temperature and Pressure Logger Specifications
| Data Storage Capacity |
21,500 Samples (1 Channel: Temperature or Pressure)
10,750 Samples (2 Channels) |
| Sampling Rate |
1 Second to 18 Hours |
| Alarms |
Over and under alarms for pressure and temperature |
| Temperature Range External Sensor |
-40°F to 150°F (-40°C to 65°C) |
| Temperature Accuracy |
±1°F(0.5°C) |
| Temperature Resolution |
0.1°F (0.05°C) |
| Time Accuracy |
±100ppm @ 75°F |
| Pressure Range |
0 - 500 psi |
| Pressure Accuracy |
±3 psi |
| Pressure Resolution |
0.15 psi |
| Pressure Connection |
1/4" NPT pressure sensor connection |
| Power Source |
9V Alkaline or Lithium Battery (Sold Separately) |
| Battery Life |
19 months using alkaline battery
38 months using lithium battery |
| Operating Temperature Range |
9V Alkaline: 32°F to 140°F (0° C to 60° C)
9V Lithium: -40°F to 150°F (-40° C to 65° C) |
| Operating Humidity Range |
0% to 95% RH, non-condensing |
| Storage Temperature |
-40°F to 170°F (-40°C to 77°C) |
| Calibration |
Single point offset calibration available through
software. Pressure is factory calibrated. |
| Dimensions |
3" x 2.5" x 1.1" (7.6cm x 6.4cm x 2.8cm) |
| Weight |
2.5oz (71g) |
Applications
- Compressor run cycles
- Pneumatic air lines
- Superheat measurements
- Testing new installations
- Tracking intermittent problems
MicroDAQ.com accepts Visa, MasterCard and American Express

All prices are in U.S. Dollars.
No U.S. sales tax
|